Automated flagging influences the inconsistency and bias of band cell and atypical lymphocyte morphological differentials

Verfasser / Beitragende:
[Wim van der Meer, Colin Stephen Scott, Marinus H. de Keijzer]
Ort, Verlag, Jahr:
2004
Enthalten in:
Clinical Chemistry and Laboratory Medicine, 42/4(2004-04-05), 371-377
Format:
Artikel (online)
ID: 378890557
LEADER caa a22 4500
001 378890557
003 CHVBK
005 20180305123453.0
007 cr unu---uuuuu
008 161128e20040405xx s 000 0 eng
024 7 0 |a 10.1515/CCLM.2004.066  |2 doi 
035 |a (NATIONALLICENCE)gruyter-10.1515/CCLM.2004.066 
245 0 0 |a Automated flagging influences the inconsistency and bias of band cell and atypical lymphocyte morphological differentials  |h [Elektronische Daten]  |c [Wim van der Meer, Colin Stephen Scott, Marinus H. de Keijzer] 
520 3 |a This study evaluated inter- and intra-observer variabilities of band cell and atypical lymphocyte differentials and the influence of instrument flagging information on resulting microscopic differentials. Five stained slides with a range of band cell counts and five with variable numbers of atypical lymphocytes were sent for morphological review by 30 technicians. No supplementary full blood cell count information was provided. Two months later, the same slides were sent, together with their corresponding analyzer reports comprising the full blood cell count, automated differentials and flags, to the same technicians. The first and second appraisals of band cells and variant lymphocytes both showed poor levels of inter-observer consistency. Observed values for all slides were very wide and suggested a high inherent predisposition to erroneous reporting practices. Analysis of category trends showed that analyzer left shift or immature granulocytes flags had no influence on observer band cell assessments as downward vs. upward category revisions were evenly balanced. The findings for atypical lymphocytes were, however, somewhat different. Two slides with no flags both showed balanced category revisions, whereas two of the three slides with atypical lymphocyte flags showed clear evidence of upward category revision. The third slide with an atypical lymphocyte flag did not show any overall category trend, but six of the seven observers who in the first examination recorded atypical lymphocyte estimates of ≤30% revised their estimates upward when the slides were examined the second time. These results suggest that morphologist access to an analyzer report and flagging information is unlikely to affect the "randomness” of band cell determinations but it may induce observer bias in variant lymphocyte estimates. 
540 |a Copyright © 2004 by Walter de Gruyter GmbH & Co. KG 
690 7 |a Medical equipment & techniques  |2 nationallicence 
690 7 |a Medical diagnosis  |2 nationallicence 
690 7 |a Diseases & disorders  |2 nationallicence 
700 1 |a Meer  |D Wim van der  |4 aut 
700 1 |a Scott  |D Colin Stephen  |4 aut 
700 1 |a Keijzer  |D Marinus H. de  |4 aut 
773 0 |t Clinical Chemistry and Laboratory Medicine  |d Walter de Gruyter  |g 42/4(2004-04-05), 371-377  |x 1434-6621  |q 42:4<371  |1 2004  |2 42  |o cclm 
856 4 0 |u https://doi.org/10.1515/CCLM.2004.066  |q text/html  |z Onlinezugriff via DOI 
908 |D 1  |a research article  |2 jats 
950 |B NATIONALLICENCE  |P 856  |E 40  |u https://doi.org/10.1515/CCLM.2004.066  |q text/html  |z Onlinezugriff via DOI 
950 |B NATIONALLICENCE  |P 700  |E 1-  |a Meer  |D Wim van der  |4 aut 
950 |B NATIONALLICENCE  |P 700  |E 1-  |a Scott  |D Colin Stephen  |4 aut 
950 |B NATIONALLICENCE  |P 700  |E 1-  |a Keijzer  |D Marinus H. de  |4 aut 
950 |B NATIONALLICENCE  |P 773  |E 0-  |t Clinical Chemistry and Laboratory Medicine  |d Walter de Gruyter  |g 42/4(2004-04-05), 371-377  |x 1434-6621  |q 42:4<371  |1 2004  |2 42  |o cclm 
900 7 |b CC0  |u http://creativecommons.org/publicdomain/zero/1.0  |2 nationallicence 
898 |a BK010053  |b XK010053  |c XK010000 
949 |B NATIONALLICENCE  |F NATIONALLICENCE  |b NL-gruyter